Determination of Iodate in Iodized Salt by Reversed-Phase High-Performance Liquid Chromatography with UV Detection.

Item request has been placed! ×
Item request cannot be made. ×
loading   Processing Request
  • Additional Information
    • Abstract:
      A method for the determination of iodate was developed by reversed-phase high-performance liquid chromatography with UV detection. Iodate was converted to iodine, which was separated from the matrix using a reversed-phase Ultrasphere C18 column (250 × 4.6 mm, 5 μm) with methanol-1 mmol L−1 H3PO4 (20:80, v/ v) as mobile phase at 1.00 mL min−1 and UV detection at 224 nm. The calibration graph was linear from 0.05 μg mL−1 to 5.00μg mL−1 for iodine with a correlation coefficient of 0.9994 ( n=7). The detection limit was 0.01 μg mL−1. The method was successfully applied to the determination of iodate in iodized salt. The recovery was from 96% to 101% and the relative standard deviation was in the range of 1.5% to 2.9%. [ABSTRACT FROM AUTHOR]
    • Abstract:
      Copyright of Chromatographia is the property of Springer Nature and its content may not be copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written permission. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)